A-dec style on/off relieving toggle releases air when in the off position (makes a hissing sound).
Product Details:
- For use with air to activate master on/off activation, water on/off activation, and arm locks on A-dec® units.
- Only for use with 1/8″ tubing.
- Tubing clamps and gaskets included.
Installation Instructions (PDF):
- Shut off your delivery unit by depressing the buttons on the air/water syringe.
- Cut the tubing as close to the old toggle with a wire cutter and remove the toggle. Make sure to note which tubing connects to the inlet and outlet.
- Install the appropriate sized barbs with washers and tighten securely with a 1/4″ open ended wrench.
- Apply liquid soap to the tubing or dip the ends in hot water. Slip the sleeve clamps over the tubing (approximately 3/8″ from the end), which ensures the interior bevel of the sleeve clamp faces the barb.
- Slip tubing over barbs and lock tubing by using a pliers or sleeve tool to slide the sleeve clamps over tubing and barb.
- Mount to the delivery unit using provided nut.
- Finally, turn the unit on in order to test for leaks and proper function.